Overview
Note: The
Auto and
Ext options are not available in Basic Balance
mode.
Procedure
- From the Balance main menu, press F3 Job Setup.
- In Basic Balance mode, press F12 Sensor Setup from the Job Setup screen. In Advanced Balance mode, press .
- Press F5 Mux Enabled.
-
Use the up and down arrow keys to select an option.
Option Description Off The analyzer does not use the CSI 2140 Four-Channel Input Adapter or external multiplexer. The analyzer acquires all data through the same channel, and you must manually start and accept each measurement. On The CSI 2140 Four-Channel Input Adapter controls software that switches between channels. The analyzer acquires all data on channels A to D, and you must specify the input channel for each measurement point. You must manually start and accept each measurement reading. The default is On. Auto The CSI 2140 Four-Channel Input Adapter controls software switching between channels. The analyzer acquires all data on channels A to D, and you must specify the input channel for each measurement point. After you start the first reading, the analyzer acquires and accepts each measurement when the value is good for three consecutive averages (stable). The analyzer advances to the next step. Ext An external multiplexer is used. The analyzer acquires all data on channels 1 to 8, and you must specify the input channel for each measurement. You must manually start and accept each measurement reading. - Press Enter to select the option.
- Press Enter.