Set the number of inputs in ODS/Modal

Overview

After you collect data, you cannot change the number of inputs for the job. You can modify the sensors.

Procedure

  1. From the ODS/Modal main menu, press F8 Job Manager > F4 Edit This Job > F12 Input Setup > F1 Select Input.
    Note: After you collect data, the F12 Input Setup key is no longer available.
  2. Select one or more inputs.
    Option Description
    F2 Select/Unselect Input A Enable or disable data collection on input A.
    F3 Select/Unselect Input B Enable or disable data collection on input B.
    F4 Select/Unselect Input C Enable or disable data collection on input C.
    F5 Select/Unselect Input D Enable or disable data collection on input D.
    F8 A and B Collect input data from A and B simultaneously. If you select this option, the inputs on the left side of the screen automatically update.
    F9 EmersonTriax Collect data from inputs A, B, and C. If you select this option, the inputs on the left side of the screen automatically update. For the CSI A0643TX triaxial accelerometer, the directions are relative to the sensor: Input A = Z direction, Input B = X direction, Input C = Y direction.
    Note: When you select Emerson Triax, the following sensor options are set for the A, B, and C inputs:
    • Sensor Type = Accelerometer
    • Sensitivity = 0.1
    • Sensor Power = On
  3. Press Enter.
    The active inputs and units display at the bottom of the screen. Disabled inputs are grayed out.