Set the number of inputs in Advanced Transient

Procedure

  1. Activate a job.
  2. From the Transient main menu, press F1 Setup / Acquire Data > F12 Input Setup > F1 Select Input.
  3. Select one or more inputs.
    Option Description
    F2 Select/Unselect Input A Enable or disable data collection on input A.
    F3 Select/Unselect Input B Enable or disable data collection on input B.
    F4 Select/Unselect Input C Enable or disable data collection on input C.
    F5 Select/Unselect Input D Enable or disable data collection on input D.
    F7 A only Collect input data from input A. If you select this option, the inputs on the left side of the screen automatically update.
    F8 A and B Collect input data from A and B simultaneously. If you select this option, the inputs on the left side of the screen automatically update.
    F9 EmersonTriax Collect data from inputs A, B, and C. If you select this option, the inputs on the left side of the screen automatically update. For the CSI A0643TX triaxial accelerometer, the directions are relative to the sensor: Input A = Z direction, Input B = X direction, Input C = Y direction.
    Note: When you select Emerson Triax, the following sensor options are set for the A, B, and C inputs:
    • Sensor Type = Accelerometer
    • Sensitivity = 0.1
    • Sensor Power = On
  4. Press Enter.
    The active inputs and units display at the bottom of the screen. Disabled inputs are grayed out.